(Id117457) Defibrillator, Automatic External; General Use (External Defibrillation) In Adult And Pediatric Patients; Portable; Biphasic Wave Form; Automatic Detection Of Ventricular Fibrillation And Ventricular Tachycardia Without Pulse With Specificity Levels And Sensitivity Above 90%; Analysis Of Thoracic Impedance; Autoteste System; Ecg Recording Memory (Minimum 30 Min.) And Events (Minimum 10); Shock Application Key; Weight (With Battery) Of Maximum 04 Kg; Liquid Crystal Display Minimas Indications (Lcd); Battery Charging Level Indicator; Display For Ecg Monitoring; Speed: 25 Mm/S At Least; Cardiac Frequency Medication Of 30 To 300 Beats Per Minute, At Least; Ecg Monitoring Via Adhesives; Defibrillation: A) Energy Scale That Covers The 150 Joules Range To 200 Joules, According To The Technology Of The Biphasic Wave (Rectilinea Or Truncated Exponential), B) Pediatric Defibrillation Mode, With Recognition And Limiting Energy Automatically By Pas, C) Maximum Load Time: Up To 12 Seconds, D) Automatic Internal Discharge When Offspring Of The Equipment, E) Automatic Internal Discharge If Not There Is Shot; Defibrillator Battery: A) Useful Life Time In Waiting At Least 04 Years, B) Minimum Of 140 Maximum Load Or 4 Hours Of Monitoring Continues (Ecg); Accompanies: A) 04 (Four) Pas Pas Games Disposable Adhesive For Adult Use With Minimum 18 Months From Delivery Date, B) 02 (Two) Disposable Pass Games For Minimum Pediatic Use Of 18 Months A From The Date Of Delivery, C) 01 (One) Bag Or Briefcase For Transportation, With Foca., (Id124837) Multiparametic Monitor, Application: Adult Use, Pediatric And Neonatal; Presentation: Lcd/Led Min 14Pol, Touchscreen, Min: 1024X768; Modular; Monitoring: Ecg, Spo2, Resp, Temp, Non -Invasive Pressure (Pni), Invasive Arterial Pressure (Pi), Cardiac Debit (Co) And Etco2 (Capnography); Simultanea Min View: 6 (Six) Curves, With Numeric Values; Sound And Visual Alarms System, With Max And Min Boundaries Programable; Automatic Reconfiguration Screen; Record Data Patient; Thermal Register; Possibility Central Connection Monitoring; Hl7; Sw In Portuguese; Ecg: Monitoring: 07 Derivations (I, Ii, Iii, Avf, Avr, Avl, V); Freq. Cardiaca (Fc): 15250Bpm; Analyze Segment St: All Derivations; Analysis Of Fatal Arrhythmias In Two Simultaneous Channels At Least (Optional); Arrhythmias (Asystolism, Ventricular Fibrillation/Ventricular Tachycardia, Biigemism, Among Others); Automatic Detection Of Pacemaker; Respiration: Freq Respiratory (Fr) Min: 6150Rpm; Adjust Alarm Apnea. Pulse Oximetry (Sp2): Technology: Low Perfusion; Present Pleismographic Curve; Sto 2: 1100%; Pulse Frequency (Fp): 40250Bpm; Pni: Systolic Arterial Pressure, Diastolic And Media; Measurement Modes: Manual, Automatic And Stat; Measurement: 30270Mmhg; Temperature: 02 (Two) Channels; Measurement: Superficial And Esophagic Sensor; Temp Rang: 2545 ° C; Capnography (Etco2): Sidestream Or Mainstream Technology, Sampling: 50Ml/Min, Co2 Drug: 099Mmhg; Pi: 02 Channels, Pi Medication: 30 To 300Mmhg, Precise: ± 1Mmhg Or ± 3%, Whatever Is The Highest; Have Physiological And Technical Alarm; Measure Pressure: Pulmonary Artery, Right Atrium, Central Venous, Left, Intracranial Atrium; Arterial, P1 And P2; Cardiac Debt (Co): 0, 120 L/Min; Variation Vpp (Or Delta Pp) Variation; Specific Cva Screen
|