Table / Bench Oximeter - Equipment Used For Measuring, Indication Of Spo2, Heart Rate, Pulse Strength, Plethysmographic Wave And Trend Tables, Through A Non-Invasive Sensor. Power Supply With Internal Bivolt Or 220V, 60 Hz, Usb 2.0 Or Higher, And Ethernet Input. Rechargeable Battery With Minimum 4H Duration With Equipment On, Spo2 Measurement Range: 70% Or Less To 100%, With Accuracy Of 3% Or Less; Pulse Rate: 30Bpm Or Less Than 240Bpm Or More Accuracy Of 5 Bpm. Must Operate In Humidity From 15% To 93% Without Condensation. It Must Weigh Between 1 Kg And 2 Kg And Come With 2 Clip-Type Sensors For Reusable Adult Patients, 1 Y-Type Or Foot Sensor For Reusable Neonatal Patients. The Equipment Must Work With Nellcor, Mindray Or Masimo Technology, As Well As Its Sensors. Non-Condensing Operating Humidity From 15% To 93%. Have Active Registration With Anvisa And Technical Assistance In The State Of Rio Grande Do Sul, Meet The Abnt Nbr Iec 60601-1 Standard And 1-Year Warranty, Instruction Manual In Portuguese, Portable Pulse Oximeter - Equipment Used For Measurement, Through A Non-Invasive Sensor. It Must Have A Rechargeable Battery With A Minimum Duration Of 10 Hours With The Equipment Turned On, Have A Charging Base With A Charging Cable And Be 220V Or Bivolt, With An Accuracy Of 3%; Pulse Rate: 30Bpm Or Less Than 240Bpm Or More Accuracy Of 5 Bpm. Must Operate In Humidity From 30% To 93% Without Condensation. It Must Weigh A Maximum Of 300G And Come With: Protective Cover With Hand Strap, 02 Clip-Type Sensors For Reusable Adult Patients, Mindray Or Masimo, As Well As Their Sensors. Non-Condensing Operating Humidity From 30% To 95%. Have Active Registration With Anvisa And Technical Assistance In The State Of Rio Grande Do Sul, Vital Signs Monitor Containing The Parameters: Heart Rate, Oxygen Saturation (Spo2), Non-Invasive Pressure (Nip) Systolic, Diastolic And Mean And Temperature Used In Patients Adults, Pediatrics And Neonatal. Screen, Minimum 7/ (640X480 Pixels), Touch Sensitive (Touchscreen), With Minimum Degree Of Protection Ipx0. Have Connectivity To Send Data To The Electronic Medical Record Via A Wired Or Wireless Network (Ethernet Or Wi-Fi Standard). Non-Invasive Adult Systolic Diastolic Pressure From 40Mmhg Or Less To 215Mmhg Or Greater With A Maximum Variation Of 5Mmhg. Infrared Or Axillary Thermometer To Measure Temperature With Amplitude With The Ability To Measure Temperatures From 34ºc Or Lower To 42ºc Or Higher. Measure Adult Saturation At 70% Or Less Than 100% With Maximum Variation Of 3% With Nellcor, Mindray Or Masimo Technology. Heart Rate From 25Bpm Or Less To 240Bpm Or Greater. 220V Power Supply. Have Active Registration With Anvisa And Technical Assistance In The State Of Rio Grande Do Sul, Instruction Manual In Portuguese. Training For Nursing Staff During Business Hours In 2 Days In The Morning And Afternoon Shifts. It Must Come With A Power Cord, A Battery With A Minimum Duration Of 3 Hours Of Use, 2 Adult Oximetry Sensors And 1 Neonatal Type Y Or Foot Sensor With Nellcor Technology, With An Extender If Necessary, 2 Adult Cuffs And 1 Neonatal Cuff With An Extender If Necessary And Car With Casters For The Monitor To Stay Fixed., 11/11/2024
|