Establishment Of Bsl-3 Laboratory On Turnkey Basis. , Supply Ahu • Capacity: 3500 Cfm • 25 Mm , • Double Skin Ahu Complete With Mixing Box, Cooling Coil, And Fan, Volume Control Dampers For Supply, Fresh Air, And Fresh Air Filter. • Static Pressure ( Mm Wg ) - 100 • Construction – 28 Mm Puf Moulded Double Skin Detachable Panels • Structural - Extruded Spl Alloy Al Profiles Filled With Puf • Panel - External Skin - 0.6 Mm Pre Plasticised Sheet • Panel - Internal Skin - 0.6 Mm Plain Gi Sheet • Insulation Material / Thickness - 25Mm Thick Puf With 45Kg / M3 Density • Fan Make & Type - Nicotra / Kruger Make Didw Backward Curve Centrifugal Fan • Fan Wheel Size & No. Of Fans - Rdh 315R #1 • Fan Speed In Rpm – 2381, Critical Fan Speed In Rpm – 3100 • Fan Outlet Velocity In Mps - 8.68 • Motor • Drive Arrangement - Taper Lock Pulleys For Fan & Motor Both • Type Of Balancing / Test Procedure - Statically & Dynamically / Amca Tested , • Type Of Vibration Isolators - Rubberized Mounts • Filters Provision - Pre Filters & Hepa Filters • Coil Type - Cooling Coil • Coil Surface Area - Sq. Ft - 6.11 • No. Of. Coil Rows - 6 • Face Velocity Across The Coil - 491 • Coil Specification - 1 / 2 " Od 27Swg Copper & 36 Swg Ai Fin At Fpi • Canvass - Srf Make Flexible Fire Retardant Canvass • Hardware - Galvanized, • Operation Wt ( Kgs ) - Aprox – 240 , Exhaust Ahu • Capacity: 4500 Cfm • Double Skin Ahu Complete With Mixing Box, Volume Control Dampers For Return, Fresh Air • Static Pressure ( Mm Wg ) - 100 • Construction - 28Mm Puf Moulded Double Skin Detachable Panels • Structural - Extruded Spl Alloy Al Profiles Filled With Puf • Panel - External Skin - 0.6 Mm Pre Plasticised Sheet • Panel - Internal Skin - 0.6 Mm Plain Gi Sheet • Insulation Material / Thickness - 25Mm Thick Puf With 45Kg / M3 Density • Fan Make & Type - Nicotra / Kruger Make Didw Backward Curve Centrifugal Fan • Fan Wheel Size & No. Of Fans - Rdh 315R #1 • Fan Speed In Rpm – 2381 • Critical Fan Speed In Rpm – 3100 • Fan Outlet Velocity In Mps - 8.68 • Motor • Drive Arrangement - Taper Lock Pulleys For Fan & Motor Both • Type Of Balancing / Test Procedure - Statically & Dynamically / Amca Tested • Type Of Vibration Isolators - Rubberized Mounts • Filters Provision - Pre Filters, Hepa Filters Hardware Etc , Ahu Control Panel , Ducting: Ventilation Ducting Will Be Made Out Of Minimum 22 Gauge Gi Sheet, Exhaust Duct From The Lab To The Ahu Will Be 22 Gauge Gi Duct. All The Ventilation Ducting Including Return Air Riser Will Be Leak Proof. Insulation Material With Nit Rile Foam / Crosslinkedpolyethylenematerial ( Aluminum Foil Insulation ) , Condensing Unit: Supply And Installation Of Condensing Unit With 8.5 Tr Capacity Including, Gas, Copper Tubing, Drier & Expansions Valves Etc , Treated Fresh Air Unit – 5 Tr , Dampers: Galvanised Steel Volume Controlled Damper, Opposed / Parallel Blade Dampers Are Used To Carry Out A Rough Air System Balance With Closer Control Being Carried Out At The Individual Grilles Or Diffusers. , Supply & Return Air Grillsmoc: Crca Powder Coated, 1.2 Mm Thick With Perforation , Supply Hepa Filter Box Type :Box Type, Media :Ultra Clean Glass Fibre Paper-Imported, Casing:M.S. Powder Coated, Retention:0.3 Micron, Efficiency:99.97%, Pressure Drop:25 Mm Of W.C. Size: 450 X 450 X 75 Mm , Bibo Filter Unit , Virus Burnout System , Ceiling Panel: Modular Ceiling: Modular Ceiling Will Be Made For Clean Room Application, Pre-Engineered 50 Mm Thick Puf Panels With Pcgi Sheets With Puf Insulation Of Minimum 38-40 Kg / M3. Both Surfaces Should Be Power Coated 0.47 Mm Thick Pcgi Sheet And Will Be Installed Along The Outer Walls, Partitions And False Ceiling To Create An Impervious Shell Which Is Fully Sealed. , Modular Wall:Modular Wall Will Be Made For Clean Room Application, Pre-Engineered 50 Mm Thick Puf Panels With Pcgi Sheets With Puf Insulation Of Minimum 38-40 Kg / M3. Both Surfaces Should Be Power Coated 0.47 Mm Thick Pcgi Sheet And Will Be Installed Along The Outer Walls, Partitions And False Ceiling To Create An Impervious Shell Which Is Fully Sealed. , Modular Wall:80 Mm Wall Panel With Return Air Riser , Covings: Extruded Aluminum Anodized R75 Clip-On Type ( Male & Female Connectors ) Covings For Entire Wall To Wall & Wall To Ceiling Joints , Modular Clean Room Door Flush Door Finishes Shall Be 50 Mm Thick 1.2Mm Thick Pcgi Sheet Suitable To Fix On 50 Mm Thick Wall Panel With Provisions For Double Glazing Glass For All Door And Hardware Like Push Plate.Puf Panels Will Be With Pcgi Sheets, Powder Coated On Both Sides And Puf Insulation Of Minimum 38-40 Kg / M3. Concealed Hardware For Fixingofdoor Frames, Door Closure, Sshinges, Ss Doorhandle, Ssball Bearing Butt Hinges, Automatic Drop Seal, Concealed Tower Bolt For The Double Door, Both Sides Lock And Key Arrangement Etc , Size: ( Single Door ) 1000 X 2100 Mm- Biosaftey Doors , Size: ( Double Door ) 1200 X 2100 Mm-Biosafty Doors , Size: ( Single Door ) 1000 X 2100 Mm-Metallic Doors , Differential Pressure Gauge Supply, Installation Of Dwyer Make Magnehelic Pressure Gauges 0 – 10 Mm For Monitoring Differential Room Pressures , Epoxy Flooring: Flooring Shall Be Of 6 Mm Of Industrial Epoxy Including Screed Compound For Adhesion With Bubble Free Perfect Smooth Finishing Completed Into Two Steps: Hardening And Smoothening. Epoxy Used For This Application Will Be Self-Leveling And Clean Room Compatible. , Epoxy Coving , Door Interlocking System , Dynamic Pass Box: Moc : Ss 304 Size: 2’ X 2’ X 2’ , Door, Interlocking System, Hepa Filter, Pre Filter, Motor Blower, Uv Light, Fluorescent Lamp Etc , Dunk Tank: Moc : Ss 304 Size: 2’ X 2’ X 2’ , Door, Interlocking System, Etc. , Electricals: For The Purpose Of Energizing The Bsl Laboratory Facility All The Electrical Fittings And Fixtures In The Laboratories Areas Shall Be Suitable For Clean Room Application And Shall Be Sealed. , Shed Work For Ahu & Ac , Uv Lamp , Led Lights , Split Ac 1 Tr, With Stabilizer, Copper Tubing Etc , Split Ac 1.5 Tr, Timer Control With Stabilizer, Copper Tubing Etc , Garment Storage Cabinet: A Garment Storage Cabinet That Can Be Locked Shall Be Provided In The Change Rooms For Each Lab Worker. The Garment Storage Cabinets Shall Be In Ss304 , Dimension: 4’ X 1.5’ X 5’ ( Moc: Crca Powder Coated ) , Elbow Operated Tap With Ss Sink Unit , Emergency Shower: Emergency Shower With Eye Wash , Laboratory Chairs , Fire Detection & Alarm System: Fire Detection And Alarm System Will Be Provided For The Entire Bsl-3 Laboratory Facility To Detect Any Eventual Case Of Fire In The Building. , Fire Extinguishers - 2 Kg , Epabx System ( Intercom Facility ) : Telephone Instrument Facility To Be Provided Between Microbiologist / Staff Room , Cctv Surveillance System With 8 Dvr And Storage Capacity With Monitor And 8 Cctv Cameras In The Bsl Iii, Ante And Change Rooms , Lan Wiring And Connectivity With Sockets To Be Provided At Strategic Locations In The Bsl Facility , Uninterrupted Power Supply System: For Interlocking System, Fire Panel, Cctv Etc ( 5 Kva ) , Uninterrupted Power Supply System For Bio Safety Cabinet ( 3 Kva ) , Drain: All Liquid Drain Coming Out From The Laboratory Will Be Connected To A Single Drain With Back Flow Prevention , Bio Safety Cabinet: Class Ii Type A2 Size: 4’ X 2’ X 2’ ( W X D X H ) Moc: Crca Powder Coated , Effluent Collection Tank: 500 L Each , Decontamination Process Tank: 500 L Each , 125 Kva Diesel Generator Set Specification ( Minimum And Tentative Capacity Subject To Approval By Incharge )
|