Calcium Cassette Ca (Calcium) 300 Tests. Reagents For Cobas C311 Analyzer. Calcium Cassette Ca (Calcium) For 300 Tests. Purpose: An In Vitro Diagnostic Test For The Quantitative Determination Of Calcium In Human Serum, Plasma And Urine. Reagents And Working Solutions: R1 Capso: A) 557 Mmol/L; Nm Bapta: 2 Mmol/L, Ph 10.0; Non-Reactive Surfactant; Preservative Sr Edta: 7.5 Mmol/L; Ph 7.3; Non-Reactive Surfactant; Preservative Storage And Transportation Conditions: Store In The Refrigerator. Shelf Life At 2 8 °C See Expiry Date On Cobas C Kit Label. When Used On Board The Analyzer At 10 - 15 °C 12 Weeks. Transportation Must Be In Thermal Containers With Cold Elements At A Temperature Of 2-8 °C, Do Not Expose To Freezing Or Exposure To High Temperatures. Calibration Is Linear At Two Points, Calibration Frequency Is After Changing The Lot. Measuring Range 0.20 - 5.0 Mmol/L (0.8 - 20.1 Mg/Dl). The Lower Limit Of Detection Of The Test Is 0.10 Mmol/L (0.4 Mg/Dl). Test Reproducibility Is No More Than 0.8%. The Supplier Is Required To Provide A Certificate Of Origin Of The Goods. Delivery Area: For Eu Countries. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification., Cassette Uric Acid Ua (Uric Acid) For 400 Tests. Reagents For Cobas C311 Analyzer. Cassette Uric Acid Ua (Uric Acid) For 400 Tests. Purpose: In Vitro Diagnostic Kit. Designed For Quantitative Determination Of Uric Acid Concentration In Serum, Plasma And Urine. Reagents And Working Solutions: R1 Phosphate Buffer: 0.05 Mol/L, Ph 7.8; Toos: 7 Mmol/L; Fatty Alcohol Polyglycol Ether: 4.8%; Ascorbate Oxidase (Ec 1.10.3.3; Zucchini) ≥ 83.5 Μkat/L (25 °C); Stabilizers; Preservative Sr Phosphate Buffer: 0.1 Mol/L, Ph 7.8; Potassium Hexacyanoferrate (Ii): 0.3 Mmol/L; 4 Aminophenazone ≥ 3 Mmol/L; Uricase (Ec 1.7.3.3; Arthrobacter Protophormiae) ≥ 83.4 Μkat/L (25 °C); Peroxidase (Pod) (Ec 1.11.1.7; Horseradish) ≥ 50 Μkat/L (25 °C); Stabilizers; Preservative R1 Is In Position B, Sr Is In Position C. Storage And Transportation Conditions: Store In The Refrigerator. Shelf Life At 2 8 °C See Expiry Date On Cobas C Kit Label. When Used On Board The Analyzer At 10 - 15 °C 6 Weeks. Transportation Must Be In Thermal Containers With Cold Elements At A Temperature Of 2-8 °C, Do Not Expose To Freezing Or Exposure To High Temperatures. Calibration Is Linear At Two Points, Calibration Frequency Is After Changing The Lot. Measuring Range 11.9 1500 Μmol/L (0.20 25 Mg/Dl). The Lower Limit Of Detection Of The Test Is 11.9 Μmol/L (0.20 Mg/Dl). Test Reproducibility Is No More Than 1.8%. The Supplier Is Required To Provide A Certificate Of Origin Of The Goods. Delivery Area: For Eu Countries. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification., Naoh-D Basic Wash Solution (Detergent 1), 2 X 1.8 L. Reagents For Cobas C311 Analyzer. Naoh-D Basic Wash Solution (Detergent 1) 2 X 1.8 L Containers. Purpose: Naoh‑D / Basic Wash Is Used As An Alkaline Wash Solution For Reaction Cells And Sampling Probes On Roche/Hitachi Instruments. Reagents And Working Solutions: Storage And Transportation Conditions: Store At Room Temperature. The Shelf Life Of The Opened Reagent In The Refrigerated Reagent Compartment On Board The Analyzer Is 1 Week. Do Not Expose To Mechanical Stress, Freezing Or Exposure To High Temperatures. The Supplier Is Required To Provide A Certificate Of Origin Of The Goods. Delivery Area: For Eu Countries. Remaining Shelf Life Of At Least 60%. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification., Transferrin Cassette Trsf (Transferrin) For 100 Tests. Reagents For Cobas C311 Analyzer. Transferrin Cassette Trsf (Transferrin) For 100 Tests. Purpose: In Vitro Test For The Quantitative Determination Of Transferrin In Human Urine On Roche/Hitachi Cobas C Analyzers. Reagents - Working Solutions: R1 Phosphate Buffer: 55 Mmol/L, Ph 7.2; Nacl: 25 Mmol/L; Polyethylene Glycol: 5%; Preservative R2 Antibodies Against Human (Rabbit) Transferrin: Titer Dependent; Nacl: 100 Mmol/L; Preservative Ready To Use. Storage And Stability Of Trsf2. Shelf Life At 2 - 8 °C: See Expiration Date On Cobas C Cassette Label. Shelf Life Of The Opened Reagent In The Refrigerator On Board The Analyzer: 8 Weeks. The Supplier Is Required To Provide A Certificate Of Origin Of The Goods. Delivery Area: For Eu Countries. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification., Auxiliary Solution Naoh-D, 66 Ml. Reagents For Cobas C311 Analyzer. Auxiliary Solution Naoh-D 66 Ml. Wash Solutions For Reagent Samples And/Or Reaction Chambers For Analyzers. Purpose: Reagent And Wash Solutions For The Sample Probe Or Chamber May Be Required To Eliminate The Influence Of Other Samples Or Samples On The Result Of The Current Analysis. The Solution Maintains The Integrity Of Reagents And Samples. Reagents - Working Solutions: Naohd: Naoh 1 Mol/L (Approx. 4%); Detergent Sms: Hcl 200 Mmol/L Smpcln1: Naoh 1 Mol/L Smpcln2: Buffer; Detergent Special Cell Cleaning Solution: Naoh 3 Mol/L (Approx. 12%); Sodium Hypochlorite Solution (< 2% Active Chlorine); Added Substance. Storage And Transportation Conditions: Store At Room Temperature. The Shelf Life Of The Opened Reagent In The Refrigerated Reagent Compartment On Board The Analyzer Is 1 Week. Do Not Expose To Mechanical Stress, Freezing Or Exposure To High Temperatures. The Supplier Is Required To Provide A Certificate Of Origin Of The Goods. Delivery Area: For Eu Countries. Remaining Shelf Life Of At Least 60%. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification., Auxiliary Solution Acid Wash Solution (Detergent 2), 2 X 1.8 L. Reagents For Cobas C311 Analyzer. Acid Wash Solution (Detergent 2) 2 Containers Of 1.8 L Each. Purpose: Acid Wash Solution (Detergent 2) Is Used As An Acid Wash Solution For Reaction Cells And Sampling Probes On Roche/Hitachi Instruments. Reagents And Working Solutions: Storage And Transportation Conditions: Store At Room Temperature. The Shelf Life Of The Opened Reagent In The Refrigerated Reagent Compartment On Board The Analyzer Is 1 Week. Do Not Expose To Mechanical Stress, Freezing Or Exposure To High Temperatures. The Supplier Is Required To Provide A Certificate Of Origin Of The Goods. Delivery Area: For Eu Countries. Remaining Shelf Life Of At Least 60%. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification., Magnesium Cassette Mg (Magnesium) For 250 Tests. Reagents For Cobas C311 Analyzer. Magnesium Cassette Mg (Magnesium) For 250 Tests. Purpose: Kit For In Vitro Diagnostics. Designed For The Quantitative Determination Of Magnesium In Human Serum, Plasma And Urine On Roche/Hitachi Cobas C Analyzers. Reagents And Working Solutions: R1 Trisa /6 Buffer Solution Of Aminocaproic Acid: 500 Mmol/L, Ph 11.25; Egtc: 129 Μmol/L; Preservative R2 Xylydyl Blue: 0.28 Mmol/L; Detergent; Preservative A) Tris = Tris(Hydroxymethyl) Aminomethane R1 Is In Position B, R2 Is In Position C. Storage And Transportation Conditions: Store In The Refrigerator. Shelf Life At 2 8 °C See Expiry Date On Cobas C Kit Label. When Used On Board The Analyzer At 10 - 15 °C 6 Weeks. Transportation Must Be In Thermal Containers With Cold Elements At A Temperature Of 2-8 °C, Do Not Expose To Freezing Or Exposure To High Temperatures. Calibration Is Linear At Two Points, Calibration Frequency Is After Changing The Lot. Measuring Range 0.10 2.0 Mmol/L (0.243 4.86 Mg/Dl). The Lower Detection Limit Of The Test Is 0.10 Mmol/L (0.243 Mg/Dl). Test Reproducibility Is No More Than 0.8%. Reagents For The Cobas Analyzer With 501. The Supplier Is Required To Provide A Certificate Of Origin Of The Goods. Delivery Area: For Eu Countries. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification., Ecotergent, 60 Ml. Reagents For Cobas C311 Analyzer. Cleaning Solution: Ecotergent, Cobas C311, 60 Ml Cassettes. Purpose: Ecotergent Is An Additive To Reduce Surface Tension In The Reaction Bath Of The Cobas C Analyzer. Hitergent Is Added To The Reaction Bath. It Acts As A Surfactant And Minimizes The Formation Of Bubbles That Could Potentially Interfere With Photometer Readings. Reagents - Working Solutions: Detergent. Storage And Transportation Conditions: Store At Room Temperature. The Shelf Life Of The Opened Reagent In The Refrigerated Reagent Compartment On Board The Analyzer Is 1 Week. Do Not Expose To Mechanical Stress, Freezing Or Exposure To High Temperatures. Delivery Area: For Eu Countries. Remaining Shelf Life Of At Least 60%. If The Product Does Not Meet The Technical Specifications, The Product Will Be Replaced Within 3 Days After Notification.
|