Full Body Bone Densitometry Device Specifications With Dual Energy X-Ray Bone Density Measurement Technology (Dxa), Intended For The Assessment Of Bone Mineral Density (Bmd), Body Composition (Lean Mass, Bone Mass, Adipose Mass And Visceral Fat) And Complementary Clinical Parameters, To Support Diagnosis, Monitoring And Prevention Osteoporosis, As Well As Fracture Risk Analysis (Frax) And Studies In Orthopedics, Pediatrics, Geriatrics, Sports And Obesity, Including Supply, Installation, Warranty, Training And Technical Assistance. The Equipment Should Allow Scanning Of The Entire Body, Ap Vertebral Column, Femur (Dual Femur), Forearm, Lateral Column And Lva (Vfa & Iva), Using A Fan Beam, Ensuring High Precision And Shorter Examination Times. The Minimum Scanning Area Must Be 2020 Mm The Equipment Must Operate With An Electrical Supply Of 127–220 Vac / 400 Va, Being Compatible With Conventional Clinical Environments. The System Must Present Reproducibility Less Than 1.0% C.V., Radiation Dose Less Than 10 Mr Per Exam And Use Opx/105 – Cei X-Ray Tube, With Voltage Of 120 Kv And Current Of 25 Ma, In Addition To Filtering Equivalent To 2.5 Mm Al. The System Software Must Allow The Automatic Calculation Of Bmd, Bmc, Bmi, T-Score, Z-Score, Body Composition Analysis, Application Of Clinical Tools Such As Frax, Orthopedic And Pediatric Analysis, In Addition To A Color Map Mode For Better Visualization. The Auto-Roi Functionality Should Automatically Position The Regions Of Interest, With The Possibility Of Manual Correction Using Editing Tools (Brush And Eraser) In Cases Of Fractures, Prostheses Or Anatomical Changes. The System Must Have An International Reference Database (Mí
|