Lung Ventilator For Ventilatory Support In Adult And Pediatric Patients. (The Equipment Must Be Intended For Invasive And Non-Invasive Ventilatory Support In Adult (Including Obese) And Pediatric Patients, For Use In Hospital Environments Such As Emergency Care, Intensive Care Units And Surgical Centers. It Must Have Basic And Advanced Ventilation Modes, Including Controlled Volume (Vcv), Controlled Pressure (Pcv), Pressure Support (Psv), Simv-Vc, Simv-Pc, Cpap, And Aprv/Bilevel When Available, In Addition To Allowing Spontaneous Ventilation With Or Without Pressure Support, And Non-Invasive Ventilation (Niv). It Must Meet The Following Operational Parameters: Adjustable Tidal Volume From 50 To 2,000 Ml; Adjustable Peep From 0 To 30 Cmh₂o; Adjustable Inspiratory Pressure Up To At Least 60 Cmh₂o; Inspired Oxygen Fraction (Fio₂) From 0.2 To 5 Seconds; Programmable; Minimum Inspiratory Flow Of 180 L/Min. The Equipment Must Have Integrated Monitoring With Real-Time Display Of Airway Pressure, Graphic Curves Of Pressure X Time, Flow X Time And Volume X Time, As Well As Flow X Volume And Pressure X Volume Loops. Plateau Pressure, Average Airway Pressure, Dynamic And Static Compliance. It Must Have Visual And Audible Alarms For High/Low Pressure, High/Low Volume, Apnea, Disconnection, Power Failure And System Failure. It Must Feature A Color Screen With A Minimum Size Of 8 (Eight Inches), An Interface In Portuguese And The Possibility Of Quick Adjustments On The Main Screen. The System Must Be Microprocessed, Operating Through A Medicinal Gas Network (Compressed Air And Oxygen) And/Or Internal Flow Generation System (Turbine Or Compressor), 220V Electrical Supply, Anesthesia Apparatus With Trolley For Support In Surgical Procedures (Anesthesia Apparatus With Trolley Is Intended For The Administration Of Anesthetic Gases And Controlled Or Assisted Ventilation In Patients. Neonatal, Pediatric, Adult And Obese Patients, Being Indicated For Use In Surgical Centers, Surgical Wards And Anesthetic Recovery Rooms. The Equipment Must Enable Invasive Ventilation Safely, Including Critical And Highly Complex Patients. Storage Of Accessories. The Gas Supply System Must Include Inlets For Oxygen (O₂), Nitrous Oxide (N₂o) And Compressed Air, With Valves And Connections In The Abnt Standard. Ml/Min And Ensure A Safety System To Supply A Minimum Concentration Of 25% O₂. The Equipment Must Have Two Universal Supports For Vaporizers Compatible With Halogenated Anesthetic Agents (Isoflurane, Sevoflurane, Desflurane And Halothane), With A Mechanical Or Electronic Switching System That Prevents The Simultaneous Activation Of Two Vaporizers In The Respiratory System. It Must Be Of The Closed Circuit Type, With One-Way Valves And A Co₂ Absorber, With A Minimum Capacity Of 1 Kg Of Granulate. The Integrated Fan Must Include Ventilation Modes
|