Mobile Hemodialysis Machine; Mounted In A Cabinet On 4 Casters (2 Of Which Have Brakes), With Grounding On All Surfaces; Of Negative Pressure And Automatic Proportion; For Use In Intensive Care Unit Or Dialysis Service;Single And Double Puncture; Microprocessed; Must Carry Out Self-Test At Equipment Startup With Blocking In Case Of Operating Failures; You Must Perform Standard, Sequential Dialysis And Hemofiltration In On-Line And/Or Infused Via Saline Vials. Dialysis With Acetate And Bicarbonate, With Possibility Of Operation With Powdered Bicarbonate; Servo Controlled Proportion System; Continuous Heparin Infusion Pump, Programmable, For Use With Commercial Syringes; For Operation With Dialyzers Of Any Type, Low And High Flow; Connectors Or Adapters That Allow Use With Arterial, Venous Lines And Universal Dialyzers Of Pediatric And Adult Sizes; Or Dedicated Lines, Ultrafiltration (Covering At Minimum): 0.1 To 2.0 L/H; Dialysate Temperature (Covering At Minimum): 35 To 39 C; Automatic Control Of Transmembrane Pressure And Dialysis Liquid Proportion; Dialysate Flow (Minimum): 600 Ml/Min; Blood Pump (Covering At Minimum): 50 To 500 Ml/Min; Equipped With An Automatic Disinfection System; Heat Or Chemical Disinfection System; Monitoring: Temperature, Dialysate Pressure (Transmembrane), Conductivity, Dialysate Flow Rate, Arterial And Venous Pressure Of The Extracorporeal System; Must Carry Out Kt/V Calculation Online; Dialyzer Rupture Detector, Bubble Detector, Air Detector With Venous Line Blocking (In Case Of Failure), Protection Against Gas Embolism; Protection Against Operation In Dialysis Mode When In Disinfection Mode And From Disinfection When In Dialysis Mode; Alarms: Audible And Visual For: - Blood Leakage; - Bubble Detection; - Abnormal Conductivity; - Abnormal Temperature; - Arterial And Venous Pressure; - Non-Invasive Blood Pressure; - Lack Of Energy; - Lack, Electric Scalpel General Purpose Electrosurgical Equipment With Power Of Up To 300 W, Intended For Carrying Out Surgical Procedures With Tissue Cutting And Coagulation Through Electrosurgery, Serving Various Specialties, Such As Endoscopy, Gynecology, Urology And Microsurgery In General. It Is An Electronic, Microprocessed And Cf Type Equipment, Designed To Offer Precision, Safety And Versatility In The Most Demanding Surgical Environments. Its Structure Comprises A Class I Cabinet, With Ipx1 Rating For Protection Against The Ingress Of Liquids Into The Scalpel And Ipx7 For The Pedal, Ensuring Adequate Resistance According To Exposure Levels In Surgical Centers. Available Power Varies Depending On Operating Mode. In Monopolar Mode, It Allows Cutting With Adjustment Up To 300 W, Coagulation Up To 120 W And Mixing (Blend) Up To 200 W, All With Increments Of 1 W. In Bipolar Mode, It Offers Adjustable Cutting Up To 100 W And Coagulation Up To 70 W. The Equipment Features Several Technological Features, Such As A Continuous Sine Wave Form Specific For Bipolar Cutting, In Addition To Pulsed Cutting, Ideal For Endoscopic Procedures. It Has Both Manual And Pedal Activation Control, With Two Distinct Cutting Modes: One Pure And Another Specific For Delicate Cuts. There Is Also The Possibility Of Connection With An Argon Gas Beam Coagulator, Convection Cooling System And Continuous Monitoring Of Contact Impedance
|