01 (One): Rental - Automated Biochemistry Analyzer With Ise Module (Monthly): Minimum Characteristics.- Minimum Speed 440 Tests/Hour With A Minimum Of 240 Photometric And 200 Ise- Maximum Reaction Volume Of 250Ul- Maximum Water Consumption Of 3Lt/H- Open System For Reagents- Have No Minimum 14 Reading Filters With Wavelengths Between 340Nm - 800Nm - Have An Automatic Cuvette Washer - Have A Refrigerated Tray For Samples And Reagents - Have A Clot Detection System - Have A System That Allows Emergency Sample Insertion - Have A Minimum Of 80 Reusable Reaction Cells - Capacity Minimum Of 30 Reagents In The Reagent Tray - Minimum Capacity Of 60 Samples In Primary Tubes In The Sample Tray - Have A Barcode Reader For Samples And Reagents - Have The Ability To Perform Self-Dilution - Have A Continuous Sample Loading System - Have A Built-In Ise Module With A Minimum Speed Of 200 Tests/Hour - Ise Module With Analysis Of Na, K, Cl And Ca - Possess The Capability Of Performing Bi-Directional Interface, 01 (One): 5-Part Hematology Analyzer Differential: Minimum Features: 5-Part Hematology Analyzer With Independent Basal Channel And Tri-Angle Laser Dispersion With Advanced Flow Decytometry Method. Hematology Analyzer Principles: Flow Cytometry (Fcm) + Tri-Angle Laser Scattering + Chemical Staining Method For Differentiation Wbc Impedance Method For Rbc And Plt Test. Cyanide-Free Colorimetry For Hgb Testing. Parameters: 25 Reportable Parameters: Wbc, Neu#, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%, Rbc, Hgb, Hct, Mcv, Mch, Mchc, Rdw-Sd, Rdw-Cv, Plt, Mpv, Pdw, Pct, P-Lcr, P-Lcc 6 Parameters Search: Aly#, Aly%, Lic#, Lic%, Nrbc#, Nrbc% 3 Histograms For Wbc, Rbc And Plt. Three Diff Scatter Diagrams And One Base Scatter Diagram For Differentiating Wbc Sample Mode: Whole Blood, Whole Capillary Blood And Pre-Diluted Modes Sample Volume: 20Μllinearity Range:Wbc: 0.00-300.00×109/Lrbc: 0.00-8.50×1012/Lhgb: 0-250G/Lplt: 0-3000×109/Lhct: 0.0-67.0%Repeatability:Wbc ≤2. 5% (4.0-15.0×109/L)Rbc ≤1.5% (3.5-6.0×1012/L)Hgb ≤1.5% (110-180G/L)Mcv ≤1% (70-120Fl)Plt ≤6.0% (100-149×109/L)Printing: 57Mm Thermal Printer With Multiple Formats.External Printer: Compatible With Multiple Laser/Inkjet Printers, Compatible With Multiple Formats And User-Defined Formatsyield: Up To 60 Tests Per Hourstorage: Up To 100,000 Recordsnet Weight: 28Interface: 4 Usb Ports, And 1 Communication Port: Lan Port Supports Hl7 Protocol Supports Bidirectional Lis Power Requirement: 100V-240V~, 50/60Hz, ≤200V Dimension: 364Mm(W)*492Mm(H)*451Mm(D), 01 (One): Data Analyzer Selective Ions: Minimum Characteristics: Sample Type Whole Blood, Serum, Plasma And Urine Sample Volume 100 Μl: Whole Blood, Serum And Plasma; 60 Μl: Capillary; 400 Μl: Diluted Urine (1:10)Direct Measurement Method By Ion Selective Electrode (Ise)Measurement Range Blood Urine˖: 20-200 Mmol/L Na˖: 25-1000 Mmol/Lk˖: 0, 2- 40 Mmol/L K˖: 1-500 Mmol/Lcl˖: 25- 200 Mmol/L Cl˖: 25-500 Mmol/Lexhibition Na˖: 0, 1 Mmol/L | K˖: 0.01 Mmol/L | Cl˖: 0.1 Mmol/Lanalysis Time Na˖/K˖/Cl˖55 Segment Urine Simanalysis Time (Urine) 90 Capillary Sample Simdata Storage Patients – Up To 128 Samples | Cq – Up To 20 Each (Normal, High, Low)Data Output Display 128 X 64 Pixels |Thermal Printer | Rs-232Amb Conditions
|