Procurement Of Instruments, Equipments, Appliances Under Rkvy Raftaar Project Namely Establishment Of Modern Disease Diagnosis Lab For Diagnosis Of Infectious Diseases Of Livestock, Poultry At Deptt Of Vety Microbiology, Cvsc, Ah, Rk Nagar Procurement Of Instruments, Equipments, Appliances Under Rkvy Raftaar Project Namely Establishment Of Modern Disease Diagnosis Lab For Diagnosis Of Infectious Diseases Of Livestock, Poultry At Deptt Of Vety Microbiology, Cvsc, Ah, Rk Nagar, Agartala , Class Ii A2 Biosafety Cabinet With Voltage Stabilizer • Biosafety Cabinet Should Have Following Electrical Requirements: Power ( Va ) : 1400, Power Of Blower ( W ) : Ac- L=330 W, M=465 W, H= 753 W. • Airflow Circulation Should Be 70% Down Flow, 30% Exhaust. • Main Filter Typical Efficiency Should Be Ulpa, 99.9995@0.12 ?m, Exhaust Filter Typical Efficiency Should Be Ulpa, 99.9995@0.12 ?m, Filters Should Be Of Reputed Brand. • Down Flow Velocity ( M / S ) Should Be 0.33, Inflow Velocity ( M / S ) Should Be 0.55. • Fluorescent Lamp Intensity ( Lux ) Should Be =900. • Net / Gross Weight Should Be ( Approx. ) : 270-293 Kg Or 595.3-646 Lbs. • Interior Dimensions Should Be 920 Mm×620 Mm×650 Mm / 36.2 Inches×24.4 Inches×25.6 Inches ( W×D×H ) . • Exterior Dimensions Should Be 1080 Mm×845 Mm×2160 Mm / 42.5 Inches×33.3 Inches×85 Inches ( W×D×H ) . • Should Have 680 To 900 Mm Adjustable Height. • Should Have Sound And Flash Alarm. • Should Comply With Cfda, Ce, En12469 Certification. • One Suitable Voltage Stabilizer Has To Be Supplied For Running The Biosafety Cabinet • Bidder Should Have A Service Center At Eastern Region. • System Should Have Minimum 1 Year Onsite Warranty. Make- Haier / Thermo Fisher Scientific / Bio-Rad , Gradient Thermal Cycler With All Accessories: • Should Have A Sample Capacity Of 96×0.2 Ml Tubes, 0.2 Ml Tube Strips Or Universal Or Standard 1×96-Well Plate Of 8×12 Format With Six Or More Peltier Heating And Cooling. • System Which Is Not Capable Of Taking Plates Will Not Be Considered. • Should Have True Gradient Capability With Dynamic Ramping Technology, Other Technologies Apart From Gradient Will Not Be Considered. • Should Have The Feature Of Dynamic Ramping With Identical Hold Times For All The 8 Rows Of Gradient. • Should Have A Temperature Differential Range Of 1-25°C Across The Rows. • Shouldhaveintuitive5.7 Inches ( 14.5Cm ) Touchscreen Interfacewhichcandisplays Graphics In High Resolution For Easy Programming. • The Touch Screen Should Be Responsive For Both Gloved And Ungloved Fingers. • Should Be Capable Of Running Reaction Volumes From 1-100 ?l. • Should Have A Maximum Ramp Rate Of 4°C / Second. • Should Have A Temperature Range Of 4-100°C. • Should Have A Gradient Range Of 30-100°C. • Should Have A Temperature Accuracy And Uniformity Of ±0.5°C. • Should Have A Memory Of >500 Programs With Further Expansion Through A Usb Flash Drive For Transfer Of Files. • Should Have Block And Calculated Temperature Control Modes. • The Software Should Be Capable Of Exporting Run Logs And System Error Logs. • Should Have Quick Boot Up Time Of Not More Than 1 Min. • Should Be Quiet In Operation. • Systemshouldhavebuiltinlibraryofstandardprotocolsforlongpcr, Fastpcr, Reverse Transcription Pcr Etc. • Should Have The Feature Of “Instant Incubation” To Keep Samples At Constant Temp For Ligation And Restriction Digests. • Should Have Power Save Mode. • Shouldbecompatiblewithallkindofplasticconsumablesandreagentsspecially Reusable Sealing Mats. • The Vendor Supplying The Instrument Should Also Have The Capability Of Supplying Cdna Synthesis Kit, Hot Start Taq Polymerase, Plastic Ware, And Horizontal Electrophoresis System With Power Pack From The Same Principal Company • One 1Kva Online Ups For Running The Gradient Thermal Cycler Has To Be Supplied. • System Should Have Minimum 1 Year On-Site Warranty. Make- Bio-Rad / Eppendorf / Applied Biosystems Gradient Thermal Cycler With All Accessories— • Should Have A Sample Capacity Of 96×0.2 Ml Tubes, 0.2 Ml Tube Strips Or Universal Or Standard 1×96-Well Plate Of 8×12 Format With Six Or More Peltier Heating And Cooling. • System Which Is Not Capable Of Taking Plates Will Not Be Considered. • Should Have True Gradient Capability With Dynamic Ramping Technology, Other Technologies Apart From Gradient Will Not Be Considered. • Should Have The Feature Of Dynamic Ramping With Identical Hold Times For All The 8 Rows Of Gradient. • Should Have A Temperature Differential Range Of 1-25°C Across The Rows. • Shouldhaveintuitive5.7 Inches ( 14.5Cm ) Touchscreen Interfacewhichcandisplays Graphics In High Resolution For Easy Programming. • The Touch Screen Should Be Responsive For Both Gloved And Ungloved Fingers. • Should Be Capable Of Running Reaction Volumes From 1-100 ?l. • Should Have A Maximum Ramp Rate Of 4°C / Second. • Should Have A Temperature Range Of 4-100°C. • Should Have A Gradient Range Of 30-100°C. • Should Have A Temperature Accuracy And Uniformity Of ±0.5°C. • Should Have A Memory Of >500 Programs With Further Expansion Through A Usb Flash Drive For Transfer Of Files. • Should Have Block And Calculated Temperature Control Modes. • The Software Should Be Capable Of Exporting Run Logs And System Error Logs. • Should Have Quick Boot Up Time Of Not More Than 1 Min. • Should Be Quiet In Operation. • Systemshouldhavebuiltinlibraryofstandardprotocolsforlongpcr, Fastpcr, Reverse Transcription Pcr Etc. • Should Have The Feature Of “Instant Incubation” To Keep Samples At Constant Temp For Ligation And Restriction Digests. • Should Have Power Save Mode. • Shouldbecompatiblewithallkindofplasticconsumablesandreagentsspecially Reusable Sealing Mats. • The Vendor Supplying The Instrument Should Also Have The Capability Of Supplying Cdna Synthesis Kit, Hot Start Taq Polymerase, Plastic Ware, And Horizontal Electrophoresis System With Power Pack From The Same Principal Company • One 1Kva Online Ups For Running The Gradient Thermal Cycler Has To Be Supplied. • System Should Have Minimum 1 Year On-Site Warranty. Make- Bio-Rad / Eppendorf / Applied Biosystems , Horizontal Gel Electrophoresis Apparatus With Power Supply- • Horizontal Gel Electrophoresis System Should Be Able To Run The Gel Size Of 7 X 10 Cm & 7 X 7 Cm, And 7 X 10 Cm • The Gel Tray Should Be Supplied Along With The Gel Tank With Safety Lid. • The Supplied Gel Trays Should Be Uv Proof And The Trays Can Be Directly Kept On The Uv Transilluminator And Should Have An Integrated Fluorescent Ruler In The Tray. • The Ruler Should Get Illuminated On Exposure To Uv Light For Easy And Safe Calculation Of The Band Movements. • The System Should Include Tape Free Gel Casting Module For Leak Free Operations. • The System Should Include Two 1.5 Mm Combs, 8 & 15 Well Fixed Height Combs Each. • The System Should Have The Option For Adjustable Height Combs With Comb Holders. • Migration Rate Of Bromophenol Blue Dye Should Be Similar To 4.5Cm / Hr ( At 75 V ) . • The System Should Have A Lid With The Safety Banana Jacks, Which Breaks The Circuit When The Lid Is Running. • The System Should Be Provided With A Bubble Leveler For Even Gel Casting. • The Gel Caster Should Have 3 Height Adjustable Screws For Balancing The Uneven Platforms For Uniform Gel Casting. • The Electrodes Should Be Colour Coded To Remove The Confusion Of Wrong Orientation. • The Lid Should Have A Safety Option So That The Lid Cannot Be Closed In The Wrong Orientation. • The Lid Should Have A Integrated Cables To Connect It To The Power Pack Directly. • The System Should Be Capable To Run Precast Ready Agarose Gels And Hand Cast Gels. • Specification For Power Supply Unit: • The Power Supply Unit Should Be Programmable Capable To Operate Four Electrophoresis Units Simultaneously For Four Identical Runs With Graphic Led Display. • The Output Range Should Be 10-300 V, 0.4-400 Ma, 1-75 W. • Constant Voltage, Current Or Power With Automatic Crossover • Memory Storage:9 Programs , 9 Steps, Timer Control : 99 Hr, 59 Min • It Should Come With Automatic Power Up After Power Failure, • Safety Features, No-Load Detection; Sudden Load Change Detection. • System Should Have Minimum 1 Year Warranty Make- Bio-Rad / Thermo Fisher Scientific / Thomas Scientific Horizontal Gel Electrophoresis Apparatus With Power Supply- • Horizontal Gel Electrophoresis System Should Be Able To Run The Gel Size Of 7 X 10 Cm & 7 X 7 Cm, And 7 X 10 Cm • The Gel Tray Should Be Supplied Along With The Gel Tank With Safety Lid. • The Supplied Gel Trays Should Be Uv Proof And The Trays Can Be Directly Kept On The Uv Transilluminator And Should Have An Integrated Fluorescent Ruler In The Tray. • The Ruler Should Get Illuminated On Exposure To Uv Light For Easy And Safe Calculation Of The Band Movements. • The System Should Include Tape Free Gel Casting Module For Leak Free Operations. • The System Should Include Two 1.5 Mm Combs, 8 & 15 Well Fixed Height Combs Each. • The System Should Have The Option For Adjustable Height Combs With Comb Holders. • Migration Rate Of Bromophenol Blue Dye Should Be Similar To 4.5Cm / Hr ( At 75 V ) . • The System Should Have A Lid With The Safety Banana Jacks, Which Breaks The Circuit When The Lid Is Running. • The System Should Be Provided With A Bubble Leveler For Even Gel Casting. • The Gel Caster Should Have 3 Height Adjustable Screws For Balancing The Uneven Platforms For Uniform Gel Casting. • The Electrodes Should Be Colour Coded To Remove The Confusion Of Wrong Orientation. • The Lid Should Have A Safety Option So That The Lid Cannot Be Closed In The Wrong Orientation. • The Lid Should Have A Integrated Cables To Connect It To The Power Pack Directly. • The System Should Be Capable To Run Precast Ready Agarose Gels And Hand Cast Gels. • Specification For Power Supply Unit: • The Power Supply Unit Should Be Programmable Capable To Operate Four Electrophoresis Units Simultaneously For Four Identical Runs With Graphic Led Display. • The Output Range Should Be 10-300 V, 0.4-400 Ma, 1-75 W. • Constant Voltage, Current Or Power With Automatic Crossover • Memory Storage:9 Programs , 9 Steps, Timer Control : 99 Hr, 59 Min • It Should Come With Automatic Power Up After Power Failure, • Safety Features, No-Load Detection; Sudden Load Change Detection. • System Should Have Minimum 1 Year Warranty Make- Bio-Rad / Thermo Fisher Scientific / Thomas Scientific , Gel Documentation System With Online Ups- • Should Be Able To Support Wide Range Of Applications Like Fluorescence, Colorimetry / Densitometry & Gel Documentation. • The Gel Imaging System Should Support The Following Dyes – Sybr Green, Sybr Safe, Ethidium Bromide, Stain Free Gel , Coommassie Blue, Zinc Stain, Flamingo, Oriole, Silver Stain, Coommassie Fluor Orange, Sypro Ruby, Krypton & Colorimetric Blots • The Gel Imaging System Should Feature Touch Screen Of Size 9.7 Inches Or More Which Is Multitouch Capable & Offer A Display Resolution Of 1024 X 768 Pixels. • The Gel Imaging System Should Offer Smart Tray Technology That Automatically Recognizes Your Application-Specific Tray And Adjust Imaging Parameters And Software Options Accordingly. • The Gel Imaging System Should Offer Pre-Calibrated Focus For Any Zoom Setting Or Sample Height. • The Gel Imaging System Should Have A High-Resolution Scientific Grade 16 Bit Cmos Camera, A Light-Tight Compact Darkroom & A Slide-Out Uv-Transilluminator. • Maximum Image Area Should Be 21 X 14 Cm ( W X H ) . • Pixel Density ( Gray Levels ) Should Be 65535. • Dynamic Range Of >3.5 Orders Of Magnitude Should Be Present. • Emission Filter Of 535–645 Nm Should Be Present. • Data Output Should Be 16-Bit Or 8-Bit: Scn, Tiff, Jpeg Image Files • The Gel Placement Door Should Be Drawer Type Allowing Access To Gels From Either Direction For Facilitating Easy / Clutter Free Gel Excision Applications. • The Imaging System Should Offer A Camera Resolution Greater Than 6.3 Megapixels & Carry A Pixel Size Of 2.4?m X 2.4?m. • The Instrument Should Offer Trans-Uv ( B ) And Epi White As Standard Illumination. White Light Trans Illumination And Uv / Blue Conversion Should Be Available As Optional. • Appropriate Flat Fielding Correction Should Be Automatically And Consistently Applied To Image Data For Every Application. • Specifications Of The Image Analysis Software: • Automated Lane And Band Identification, Molecular Weight Or Base Pairs Evaluation, Band Sizing, And Quantisation Based On A Reference Band Or Quantity Standards Should Be Present. • Snapshot Tool To Copy Images, Lane Profiles, And Graphs Should Be There. • Should Allow Publishing Resolution ( Dpi ) And Publishing Dimension To Be Specified With A One-Click Image Export For Publication. • Should Provide Functionality To Produce Image At User-Defined Dpi And Dimension • Should Not Require License For Registration. The Full Version Software Should Be Installable In Large Number Of Computers. Lifetime Free Upgrades Of Software & Firmware Should Be Available. • Software Should Be Mac And Pc Compatible • 16-Bit And 8-Bit Tiff Images With A One-Click Export Option Should Be Present. • Software Should Produce Customizable Reports With Data Organized As Desired, Including, Lane And Band Identification, Molecular Weight Or Base Pair Evaluation. Band Sizing And Quantification Are Based On A Reference Band Or Quantity Standards. • Software Should Offer Live Update Of Results With Any Change Of Analysis Parameters. • Local / Globalbackground Subtraction For Individual Bands • Should Comply With U.S. Fda 21 Cfr Part 11 Regulations. • Specification For Computer System: • Cpu Core I3, Ram -4 Gb, Hdd-1Tb, Usb2, Os, Win7 / 10 Professional, 64 Bit, Wide Screen Of Minimum 18 Inches. • System Should Have Minimum 1 Year Warranty • A 2 Kva Online Ups Has To Be Supplied. Make- Bio-Rad / Thermo Fisher Scientific / Cole Parmer , Uv Transilluminator- • Should Come With Dual Wave Length Range Of 254 Nm & 365 Nm. • 8 U.V Tubes Should Be Fitted. • Accuracy Should Be 50 Ng. • Should Have Large View Size Of 210 X 260 Mm. • System Should Have Minimum 1 Year Warranty Make- Laby / Tarson / Bio-Rad , Trinocular Research Microscope ( Inverted ) - • Ergonomically Designed Microscope Stand For Long Time Examination Of Specimens Without Any Fatigue And Inconvenience To The Operators. • Stand Should Have A Convenient Low Positioned Coaxial Coarse And Fine Focus Knobs And Also The Light Intensity Regulating Knob Positioned Near To The Focus Knobs. • Stand Should Have More Metallic Components For Long Lasting And All The Plastic Parts Should Be Of Skin Friendly Material. • Microscope Should Have An Integrated Trinocular Photo Tube Built-In 50 / 50 Beamsplitting”.With 45Deg Inclination And With Sidentopf Swivelling Eyepiece Tubes, With Ipd Adjustment Range From 48-75Mm. • Microscope Stand Should Have A Focusable Quadruple Revolving Nosepiece With Precision Click Stops For Para-Centric Observation Of Specimen Under Various Magnifications. • Microscope Stand Should Have A Fixed Large Specimen Stage 200X239 Mm With An Option To Enlarge The Specimen Stage By Another 100Mm To Accommodate Large Specimens. The Stage Should Have A Provision To Attach A Mechanical X-Y Specimen Guide With Universal Specimen Holders And As Well Specimen Holders For Various Culture / Specimen Plates / Dishes. • Microscope Should Have 6V / 30W Halogen Illumination Device And Should Also Have A Provision To Switchover To Long Life Warm White Light Led Illumination As Optional. • Microscope Should Have An External Power Adapter For Mains Supply Of 100-240V, 50-60Hz For Easy Exchange / Changeover Of Power Supply Without Any Service Technician Requirement. • Microscope Illumination Should Have Provision For On / Off Button With Additional Auto Power Off Function For Energy Conservation As Well For Lamp Life Saving. • Microscope Should Have An Infinity Corrected Optics Design And Must Have Anti-Fungus Treated Optics. • Microscope Should Have Bright Field And Phase Contrast Illumination Techniques With Easy Changeover Illumination Slider And Filter Slider With Interference Green Filter For Phase Contrast. • Microscope Should Have A Modular And Removable Long Working Distance 0.30 Na Condenser With Working Distance Of 72Mm. The Microscope Should Have Upgrade Provision For High Resolution 0.40 Na Condenser, 55Mm Working Distance, For High Magnification Objectives. • Infinity Corrected Long Working Distance Plan Achromatic Objectives With Magnifications10x / 0.25Ph1 ( W.D=4.4Mm ) , And 40X / 0.50Ph1 ( Wd=2.7Mm At D=1Mm ) ) With Suitable Common Single Ph Illumination Slider., Should Have Provision For Bottom Or Cover Slip Thicknesses Between 0.17 And 0.6Mm: Coverglass Cap For D=0.17-0.6Mm. • Wide Field 10X / 20Mm Fov Eyepieces With At Least One Focusable Eyepiece With ±5 Diopter Adjustment And With Soft Rubber Eyecups, Suitable For Spectacle Wearers Should Be Supplied. • Microscope Shall Be In Conforming To International Certification Like Cs, Ec Declaration Etc. And Certificates In This Regard Will Have To Be Furnished. • Vendor To Mention Details Of Their After Sales Support Available With Name And Designation. • Direct Presence Of Manufacturer In India Is Preferable. • Should Have Minimum One Year On-Site Warranty. • Vendor To Submit Point To Point Technical Compliance Statement. Make- Leica / Nikon / Carl Zeiss , Scientific Digital Colour Camera For Trinocular Research Microscope With Softwares And Drivers- Specification For Scientific Digital Colour Camera • Should Be Compatible For Carl Zeiss Primo Star Trinocular Research Microscope • Digital Imaging System Comprising Of 5 Megapixel Scientific Digital Color Cmos Camera • Basic Resolution: 2560 ( H ) X 1920 ( V ) = 5.0 Megapixel • Pixel Size:2.2 ?m X 2.2 ?m • Readout Of Sensor Sub-Regions ( “Roi” ) :Adjustable • Digitization:3 X 8 Bit / Pixel • Integration Time:100 ?s Up To 2 S • Interface ( Camera ) :Usb 3.0 Micro-B • Interface ( Pc / Board ) :Usb 3.0 Standard A • Spectral Range:Approx. 400 Nm – 650 Nm, Ir Filter • Read-Out Mode:Progressive • Should Have Warranty Of At Least 1 Year Specifications For Softwares And Drivers • Should Have Following Functions: Camera Control Function Like Auto White Balance, Auto Color Balance, Gamma, Contrast, Offset Saturation, Exposure Adjustment; Control Function Like Auto White Balance, Auto Colour Balance, Gamma, Contrast, Offset Saturation, Exposure Adjustment Etc. Capture Both Video And Still Images. Post Processing Function And Measurement Capability, Including Size, Angle, Area, Perimeter, Circle Radius Etc. • Should Support Window Systems Xp / 7 / 8 Or Higher. Make- Tuscen / Carl Zeiss / Nikon , Bod Incubator- • System Should Have Compact Footprint, Microprocessor Control Temperature Range Of 4°C To 60°C That Require High Or Low Temperature Levels • Capacity Should Be 150 Liter Or More. • System Should Be Ideal For Incubation And Storage At Or Below Ambient Temperature & Ideal For Sample And Chemical Storage, Microbiological Sample Testing, Water Pollution Testing, And Biochemical Oxygen Demand Protocols. • Should Have Adjustable Over And Under Temperature Alarm Alerts Operator Of Unexpected Temperature Deviations. • Should Have Single-Set Point Microprocessor Control For Precise Temperature Setting. • Should Have Rs232 Data Interface For Temperature Monitoring. • Should Have Automatic Defrost For Uninterrupted Long-Term Applications Below Ambient. • Should Have Compressor & Heater For Temperature Control. • Should Have Timer Function For Automatic Start Or Shut Down Of Unit. • Should Have Timer Of0 To 99:59 ( Hours : Minutes ) • Should Have Over Temperature Cut-Out And Fan Motor Overheating Protection For Additional Safety • Should Have Glass Window In Door For Undisturbed Inspection Of Samples • System Should Be Easy-To-Clean: Impact-Proof Interior With Rounded Edges • System Must Have Easy To Use Soft Touch Control Panel • System Must Be Ce Certified. • System Should Have Temp. Range 4 – 60°C. • System Must Have Temperature Deviation Over Time At 20 ºc: = ±0.3, At 37 ºc: = ±0.3 Or Less • System Must Have Temp. Deviation, Spatial At 20 °C: = ±0.8, At 37 °C: = ±0.7 • Should Be Supplied With 2 Number Of Shelves With Holding Capacity Of 15 Kg Each. • Electrical Requirements: 200W, 230V, 50Hz • Weight Should Be Approximately 50-55 Kg ( 110-121 Lb. ) • Should Have Air Circulation Fan In Ceiling Of Work Chamber. • System Should Have Refrigerant Type- R134a Or Better To Comply Complies With Eu-F-Gas Regulation No. 517 / 2014 • External Dimensions / Foot Print ( W X D X H ) Should Not Be More Than 27 X 25 X 44 Inches. • Should Have A Warranty Period Of At Least 1 Year. • One Suitable Voltage Stabilizer Has To Be Supplied Along With The System. Make- Thermo Fisher Scientific / Cole Parmer / Thomas Scientific , Micropipette Set With Micropipette Stand 3 ( Three ) Sets Of 6 ( Six ) Micropipettes Each- • Should Be Light Weight Mechanical Pipette. • Should Have Four-Digit Display With Display Window Facing User With 2 Button Operations. • Should Have Provision For The Customer To Adjust The Pipette For Better Accuracy To Use Difficult Liquids Or High-Density Liquids. Also, It Should Be Adjusted Back To Factory Setting Without The Need Calibration. • Should Have Quick Connection Clip To Remove The Lower Part Easily. • Pipette Piston Should Be Made Of Biopolymer Fotron, Should Be Highly Resistant To Heat, Acids And Alkalis, Mildew, Bleaches, Aging, Sunlight And Abrasion. • Should Have Very Low Tip Ejection Force – 3.6N • Should Have Spring Loaded Tip Cone: Tip Cone Provided With Spring To Reduce Stress For Up To 1 Ml Pipette. • Should Have Volume Adjustment With Few Turnarounds From Maximum To Min Volume. • Should Have Viable Calibration Seal To Indicate Factory Calibration Not Changed. • Without Removing Any Parts, Full Pipette Needs To Withstand Autoclave Sterilization At 121°C For 20 Min., There Should Be No Discoloration Upon Uv Irradiation. • Each Set Should Have 06 Micropipettes Of Following Volume Ranges: 0.1-2.5 ?l, 0.5-10 ?l, 2-20 ?l, 10-100 ?l, 20-200 ?l, 100-1000 ?l. • One Micropipette Stand Has To Be Supplied Along With The Micropipetteset • Pipettes Should Come With Warranty Of At Least 2 Years. Make- Eppendorf / Bio-Rad / , Upright Solid Door Deep Freezer ( -20°C ) With Stabilizer- • Should Have Capacity Of 344 Litres Or More. • Should Have One Solid Door, With Dimension Of: 24 Inches X 24 Inches X 73 Inches ( W X D X H ) . • Should Have Temperature Range Of – ( Minus ) 16 °C To – ( Minus ) 25 °C. • Rear Wheels And Locks Should Be Present. • Should Be Able To Defrost Manually. • Rated Load, Wattage Should Be 180. • It Should Have 8 ( Eight ) No. Of Baskets / Shelves. • Digital Temperature Controller Should Be Present • System Should Have Minimum 1 Year Warranty • A 2 Kva Stabilizer Of Reputed Company Has To Be Supplied. Make- Vestfrost / Thermo Fisher Scientific / Haier , Double Door Refrigerator- • Should Have Double Door With Top Freezer. • Should Be Frost Free. • Compressor Should Be Inverter Linear Compressor. • Capacity Should Be 335 Liters Or More. • Should Have Built In Stablilizer. • Should Be Multi Air Flow Type. • Body Should Be Made Of Stainless Steel. • Power Consumption Should Be 212 W. • Power Requirement Should Be Ac 100-310 V. • Dimensions Should Be 1700 Mm×685 Mm×600 Mm ( H×D×W ) . • System Should Have Minimum 1 Year Warranty. Make- Lg / Samsung / Godrej , Dry Bath- • System Should Have Digital Control. • Should Have Digital Display Of Time & Temperature. • Should Have Built-In Temperature Sensing Probe For Outstanding Temperature Accuracy And Control. • Controller Type Should Be Pid Digital. • Temperature Range Should Be Ambient + 5°C – 130°C ( Ambient At 25°C ) . • Timer Range Should Be 0-99:59 Min Or Continuous. • Dimensions Should Be: 288 × 200 × 100 Mm / 11.3 X 7.9 X 3.9 Inches ( L X W X H ) . • Should Have Precise Temperature Control With Pid Circuit. • Timer Should Allow The User To Accurately Monitor The Heating Time. • Should Have Powder-Coated Steel Body Construction, Interchangeable Aluminum Alloy. • Block Accommodation Should Be 28 X 1.5Ml. • System Should Be Ce Certified. • Should Come With At Least Two Years’ Warranty. Make-Thermo Fisher Scientific / Cole Parmer / Bio-Rad , Tabletop Refrigerated Micro Centrifuge- • Refrigerated Microcentrifuge Should Come With 24 X 1.5 / 2.0 Ml Rotor With Clickseal Biocontainment Lid. • Dimensions Should Be 12.9 X 11.6 X 17.5 Inches / 33.0 X 29.5 X 44.5 Cm ( H×W×D ) . • Should Have Acceleration Of Up To 21, 000 X G In As Little As 12 Seconds. • Should Be Corrosion Resistance With Lightweight Engineered Polymer Rotors. • The Rotor Should Be Able To Run 24 Micro Centrifuge Tubes In A Single Row, From 1.5 To 2.0 Ml Tubes To Minipreps And Spin Columns. • Run Time Should Be 1 To 99 Min ( 1Min Increments ) . • Should Comply Iec 61010-1, Iec 61010-2-020 Standards. • Should Have Ergonomic One-Click Rotor Opening And Closing. • Should Have Intuitive Controls, Easy-To-Read Displays And Fast One-Click Centrifuge Lid Closure. • Should Be Sleek, Lightweight, Space-Saving Design. • Should Have Max. Speed Of 14, 800 Rpm / Max. Rcf 21, 100 X G. • Should Have Following Electricity Requirements: Hertz: 50 / 60 Hz, Voltage: 230 V • Should Have Noise Level Of 50 Dba • Should Have Temperature Range Of – ( Minus ) 9°C To + ( Plus ) 40°C. • Should Comply With Csa, Ce, Ivd Certifications. • Should Come With At Least 2 Years Warranty. Make- Thermo Fisher Scientific / Cole Parmer / Eppendorf , Microvolume Spectrophotometer- • Wavelength Range Should Be 260 & 280 Nm. • Minimum Sample Size Should Be 1 ?l. • Pathlength Should Be 0.05 Mm. • Light Source Should Be Light Emitting Diodes • Detector Type Should Be Silicon Photodiode. • Spectral Resolution Should Be<8.0 Nm. • Absorbance Accuracy Should Be 3% ( At 1.05 Abs At 260 Nm ) . • Absorbance Range Should Be 0 – 30 Abs. • Detection Limit Should Be 4 Ng / ?l ( Dsdna ) , 0.12 Mg / Ml ( Bsa ) . • Maximum Concentration Should Be 1, 500 Ng / ?l ( Dsdna ) , 45 Mg / Ml ( Bsa ) . • Measurement Time Should Be < 5 Seconds. • Sample Pedestal Material Of Construction Should Be 303 Stainless Steel And Quartz Fiber. • A 1 Kva Online Ups Has To Be Provided Along With The System. • System Should Have Minimum 1 Year Warranty. Make- Thermo Fisher Scientific / Shimadzu / Denovix , Shaker Incubator- • Should Have 6 Cubic Feet Cap. • Should Be Able To Hold 250 Ml X 12 Flasks. • Inner Side Should Be Made Of Ss & Outer Of Ms Sheet. • Temp. Range Should Be 5°C To 60°C. • Should Be Controllable By Digital Indicator / Controller, Accuracy Should Be Of + / - 0.5°C. • 1 No. Of Shaker Should Be Present At The Bottom Of The Working Chamber With Speed Control, • Shaking Speed Should Be 50 - 250 Rpm. • Should Be Provided With A Motorized Fan Blower. • Electricity Requirement Should Be 230 V Ac. • Should Have Approval Of Ce & Iso. • System Should Have Minimum 1 Year Warranty. Make- Ovfu / Cole Parmer / Thermo Fisher Scientific , Mini Spin Centrifuge- • Should Make Less Noise For Operation • Should Have A Independent On / Off Switch • Should Have Fixed Speed Of 6000 Rmp / 2000G • Should Have 8 Places Rotor For 8 × 200 ?l Tube Strips • System Should Have Minimum 1 Year Warranty Make- Tarson / Eppendorf / Thermo Fisher Scientific , Air Conditioner With All Accessories- • Should Have 1 Indoor Unit, 1 Outdoor Unit, Inter Connecting Pipe, 1 Remote, 2 Batteries, Inter Connecting Wire. • Should Be Split Type. • Capacity Should Be 1.5 Ton. • Cooling Capacity Should Be 5000W. • Compressor Should Be Dual Rotary. • Should Have Dehumidification And Remote Control Facility. • Should Have Copper Condenser Coil. • Indoor Unit Dimension Should Be Approximately 83Cm×30Cm×18Cm ( W×H×D ) . • Indoor Unit Weight Should Be 8-10 Kg. • Outdoor Unit Dimension Should Be Approximately 77Cm×54Cm×28Cm ( W×H×D ) . • Outdoor Unit Weight Should Be 31-33 Kg • Indoor Temperature Indicator Should Be Present. • Cooling Coverage Area Should Be More Than 170 Sq.Ft. • Auto Air Swing Facility Should Be There • Should Have Silent Operation. • Power Requirement Should Be Ac 225-235 V. • Power Consumption Should Be 1045-1055 W. • Should Have At Least 10 Years On-Site Warranty On Compressor And 5 Years On Pcb. • The Item Has To Be Installed By The Supplier Free Of Cost. Make- Lg / Voltas / Whirlpool , Desktop With All Accessories And Printer With Cartridges- Specification For Desktop • Should Have Intel Core I3 Processor ( 10Th Gen ) • Operating System Should Be Windows 7 Or Above • Display Size Should Be = 18 Inches • Display Resolution Should Be 1920 × 1080 Pixel • Should Have Minimum 1 Tb Hard Drive. • Ram Should Be 4Gb Ddr4 • Should Come With Usb, Bluetooth And Headphone Jacks. • Realtek Wi-Fi 5 And Ethernet And 10 / 100 / 1000 Gigabit Ethernet Lan Should Be Present. • Should Come With Mouse And Key Board Of Same Brand. • Should Have At Least 1 Year On-Site Warranty. • One Suitable Ups Of Reputed Brand Has To Be Supplied. Specifications For Printer • Should Be Able To Print, Copy And Scan, Printing Method Should Be Laser, Type Should Be Multi-Function, Printing Output Should Be Monochrome • Refill Typeshould Be Toner Cartridge • Print Speed Should Be Minimum 20 Ppm, Auto Document Feeder Should Be Present • 2 Line Lcd Display Should Be Present, Scan Size ( Adf ) Should Be 216×356 Mm • Scan File Format Should Be Pdf, Jpg, Tiff, Png, Bmp, Scanner Type Should Be Flatbed, Adf, Scan Technology Should Be Contact Image Sensor ( Cis ) • Scan Resolution, Hardware Should Be Up To 4800 × 4800 Dpi • Scan Resolution, Optical Should Be Up To 600 × 600 Dpi • Should Be Able To Function Using Wireless As Well As Ethernet Networking • Should Be Hi-Speed Usb 2.0 Enabled • Should Be Compatible With Following Operating System: Windows®: 7 ( 32 / 64 Bit ) , 2008 Server R2, 8 ( 32 / 64 Bit ) , 8.1 ( 32 / 64 Bit ) , 10 ( 32 / 64 Bit ) . • Should Have Following Certification And Compliance: Ul Listed; Csa ( Canada ) ; Ce Marking; Nemko; Fimko; Demko; Semko; Eastern European • At Least 1 Year Onsite Warranty Should Be There. • The Supplier Has To Supply 05 Numbers Of Extra Compatible Cartridges Along With The Printer. Make- Hp / Dell / Toshiba
|